Optimization of intrinsic uniformity Uniformity

Lieferzeit: Lieferbar innerhalb 14 Tagen

39,90 

Integral and Differential of CFOV and UFOV of Gamma Camera

ISBN: 6200311757
ISBN 13: 9786200311757
Autor: Liaqat, Maryam/Shahid, Shaukat Ali
Herausgeber: Ali Raza
Verlag: LAP LAMBERT Academic Publishing
Umfang: 88 S.
Erscheinungsdatum: 04.10.2019
Auflage: 1/2019
Format: 0.6 x 22 x 15
Gewicht: 149 g
Produktform: Kartoniert
Einband: KT
Artikelnummer: 8131345 Kategorie:

Beschreibung

The present study deals with the determination of the effects of source geometry, gamma source activity and source volume on the intrinsic uniformity. A comparison of intrinsic uniformity for different methods for preparation of point source with extrinsic uniformity was done. Statistics was applied to analyze the experimental data. Different methods were used to make best optimized spherical point source for measurement of intrinsic uniformity. For this purpose I was used cotton swab point source and syringe point source for best geometry and performed the test at 5% UFOV recommended by NEMA (1994), Elkamhawy et al (2000) and Zanzonico (2008) and also according to the technique recommended by siemens (2002) and AAPM (1987). And conclude the following results Syringe method produced a uniform flood source and the geometry of the point source retained and best results of intrinsic uniformity. Extrinsic uniformity was time deserving due to its low activity and it was not preferable for the system uniformity detection. Intrinsic uniformity results at 5 % UFOV which is recommended by NEMA (1994) and Zanzonico (2008) is strongly preferable.

Autorenporträt

I am Physicist working on Biophysics linked with uniformity of PET and SPECT cameras as well as on Dosimetry. Done my Ph.D. in Electrical Engineering in the field of Photonics which deals with developing and designing of a Diagnosis techniques of Breast Cancer. Have short but very addressing research work in Biophysics.

Das könnte Ihnen auch gefallen …